High-flow nasal cannula, Owgels brand (model OGM-70E)
Objective : Increase the concentration of oxygen levels in the body. To reduce the rate of endotracheal intubation In patients with respiratory problems
1. Details
High pressure nasal oxygenator, Owgels brand , model OGM-70E
– Size (width x length x height) 34 x 23.8 x 16.9 cm.
– Weight 3.5 kilograms
– Power used 350 watts
– Use AC power 220 V 50 Hz.
2. Feature details
– Oxygen concentration adjuster (Oxygen blender)
Oxygen levels can be adjusted from 21-100% in steps of 1 liter/minute.
There is an Oxygen Analyzer which is a type of Ultrasonic Oxygen concentration sensor to check the concentration of oxygen.
– The machine can adjust the flow rate (Flow rate).
The air flow rate can be adjusted from 2 – 70 liters / minute.
in child mode Who want to use Low flow mode can adjust the air flow rate 2 – 25 liters / minute.
– Humidity and temperature generator
The machine is divided into 2 parts: the heat generating part and the part that creates humidity
– System with automatic temperature indication numbers Able to read temperature values on the radiator (Chamber) precisely
– The temperature can be set from 31 – 37 °C.
– Humidity < 10 mg/L at temperature 31 °C
– Humidity < 10 mg/L at temperature 34 °C
– Humidity < 33 mg/L at temperature 37 °C
– It takes approximately ≤ 30 minutes to warm up the machine.
– Displayed with a digital screen Convenient to use
– Can be connected with an oxygen generator. and oxygen tank The temperature of the air from the oxygen supply must not exceed 43 °C.
– There are various warning signs on the screen with 4 functions:
Low-Higher than normal concentration alert or set
Clock symbol notification when the set time is approaching.
Temperature symbol warning When the machine detects abnormal temperatures within the system
– Notification with sound, 9 functions including:
When the machine detects that the air pipe or air pipe is not connected properly, is leaking, or is clogged, there is a problem and damage. until unable to send air to the user
Notification when found Position for connecting pipes or equipment It was connected in the wrong position. or in the wrong direction
When the machine detects Air flow rate lower than set
When the machine detects Air flow rate higher than set
When the machine detects The water level in the boiler provides all the moisture. or the water level is lower than the recommendations
When the machine detects The oxygen supply has a low concentration, low pressure, or higher than 0.60 MPa.
When the device detects a lower temperature than the user has set
When the machine detects a higher temperature than the factory has set
When the machine detects that there is a blockage in the airway This may occur for many reasons. such as a clogged mask
or deteriorated sausage, etc.
– You can choose to view data in 3 graph formats: oxygen concentration over time graph,
Graph showing air flow rate over time. and a graph showing temperature over time
3. Product safety standards
– Manufactured according to international standards ISO 13485 and received CE certification.
4. Equipment for use
– OGM-70E unit (1 unit)
– Quality certificate that has inspected the machine from the factory (1 piece)
– Manual with usage instructions (1 set)
– AC plug cable, length 1.8 meters (1 piece)
– Air filter (2 pieces)
– Nasal oxygen tube For high pressure oxygen machine (2 pieces)
– Humidifying water boiler (1 piece)
– Humidifying water boiler connector (1 piece)
– Air pipe set (1 piece)
– Low pressure oxygen gas connection cable, 1.5 meters long (1 line)
– High pressure oxygen gas connection line (1 line)
Product warranty
Product guaranteed for 2 full years
There is a service repair center. with a team of technicians providing service Ready to give advice How to use 24-hour service